5.5 Method run subwindow

■■■■■■■■■■■■■■■■■■■■■■

428

■■■■■■■■

viva 1.0

Device type

Display or selection of the instrument type. If an instrument is selected in
the Device name field, the Device type field can no longer be edited
and the instrument type belonging to the instrument is displayed.

If the not defined option is selected in the Device name field, all instru-
ment types or groups with which the command can be executed can be
selected, regardless of the instruments present in the device table.

Selection

Avantes Spectrometer

Default value

Avantes Spectrometer

5.5.2.5.3.3

MEAS Spec - Measuring parameters

Tab: Method

MEAS Spec Properties... Measuring parameters

Command name

Name of the command.

Entry

25 characters

The parameters for the measuring procedure can be defined on this tab.

Start wavelength

Lower limit of the spectrum.

Input range

100.0 - 2,000.0 nm

Default value

400.0 nm

End wavelength

Upper limit of the spectrum.

Input range

100.0 - 2,000.0 nm

Default value

1,000.0 nm

Integration time

Integration time for the acquisition of the spectrum.

Input range

0.01 - 600,000 ms

Default value

6 ms

Averaged spectra

The number of spectra that are acquired and averaged.

Input range

1 - 10,000

Default value

10

Smoothing

Number of neighboring pixels with which the value for each measurement
pixel is determined.