2 editing templates for input lines, 2 templates - output lines, 1 managing templates for output lines – Metrohm viva 1.0 Manual User Manual
Page 663
■■■■■■■■■■■■■■■■■■■■■■
6 Configuration
viva 1.0
■■■■■■■■
651
Delete
Deletes the selected template.
6.3.3.1.2
Editing templates for input lines
Dialog window: Configuration
▶ Tools ▶ Templates ▶ Input lines... ▶ Templates
for input lines
▶ New... / Properties... ▶ New template / Edit template
Signal name
Name of the template for the input signal.
Entry
25 characters
Input signal
Entry of the bit pattern for the input signal with exactly 8 characters. It is
possible to enter the characters
0 = Line inactive,
1 = Line active and
* = Any line status.
Selection
Bit pattern with 8 characters (0, 1, *) | ********
Default value
********
The input lines and bits are numbered from right to left:
Input 7 6 5 4 3 2 1 0
Bit
7 6 5 4 3 2 1 0
Example: *******1 expects an active input line 0 (1 = set).
NOTE
Input lines that are of no interest or for which no defined condition can
be predicted should also be masked with an asterisk *.
6.3.3.2
Templates - Output lines
6.3.3.2.1
Managing templates for output lines
Dialog window: Configuration
▶ Tools ▶ Templates ▶ Output lines... ▶ Templates
for output lines
Client-specific bit patterns for setting remote output signals can be
defined in the Templates for output lines dialog window and selected
with the CTRL command and in manual operation. The table with the
defined templates cannot be edited, although it can be sorted according
to the selected column in ascending or descending order by clicking on
the column title.