Metrohm viva 1.0 Manual User Manual
Page 518
5.5 Method run subwindow
■■■■■■■■■■■■■■■■■■■■■■
506
■■■■■■■■
viva 1.0
Selection
'Device name' | not defined
Default value
not defined
not defined
In this case, the device name must be assigned by the user at the start
of the method.
Device type
Display or selection of the instrument type. If an instrument is selected in
the Device name field, the Instrument type field can no longer be
edited and the instrument type belonging to the instrument is displayed.
If the not defined option is selected in the Device name field, all instru-
ment types or groups with which the command can be executed can be
selected, regardless of the instruments present in the device table.
Selection
814 USB Sample Processor | 815 Robotic USB
Sample Processor XL | 858.0010 Professional
Sample Processor | 858.0020 Professional
Sample Processor | 858.0030 Professional
Sample Processor | 919.0020 IC Autosampler
plus
Default value
858.0020 Professional Sample Processor
Target
Tower
Selection of the tower on the sample changer at which the robotic arm is
to be moved to the required position.
Selection
1 | 2
Default value
1
Swing
Selecting the target position for the robotic arm.
Selection
External position | Maximum angle | Relative
angle
Default value
External position
External position
Swinging to one of the 4 external positions available for each tower.
They are defined in the tower properties of the Sample Processor.
Maximum angle
Swinging to the maximum permissible outward angle (see Chapter
7.3.8, page 747).