5.5 Method run subwindow
■■■■■■■■■■■■■■■■■■■■■■
504
■■■■■■■■
viva 1.0
Cancel determination
The determination is canceled and the next determination of the series
is started. If an exit track is present, it will be called. No error track will
be called.
Cancel determination and series
Both the determination and the series are stopped. If an exit track is
present, it will not be called. An error track will be called.
Parameters
Shift rate
Selection of the rate at which the rack rotates.
Input range
3 - 20 °/s
Default value
20 °/s
Shift direction
Selection of the shift direction.
Selection
auto | + | –
Default value
auto
auto
The shift direction with the smallest rotation angle is selected automat-
ically.
+
Rack rotates counterclockwise.
–
Rack rotates clockwise.
Swing rate
Selects the rate of the robotic arm when moving to a rack position or a
special beaker (only with multiple row racks).
Input range
10 - 55 °/s
Default value
55 °/s
5.5.2.8.3
SWING
5.5.2.8.3.1
SWING - Overview
Dialog window: Method
▶ SWING ▶ Properties... ▶ SWING - 'Command name'
Command for swinging the robotic arm. This command cannot be
executed unless the Sample Processor is equipped with a 786 Swing
Head with robotic arm.
Instruments
This command can be executed with the following instruments: