5.6 Evaluation subwindow
■■■■■■■■■■■■■■■■■■■■■■
558
■■■■■■■■
viva 1.0
Max. width
Maximum peak width for the assignment of a peak to a substance. The
calculated peak width must be smaller than Max. width in order for the
substance to be recognized.
Input range
(Min. width) - 1.000 V
Default value
0.500 V
Min. measured quantity
Minimum measured quantity for the assignment of a peak to a substance.
The measured quantity must be greater than Min. measured quantity
in order for the substance to be recognized. Value and unit can be
entered or selected separately.
Input range
1 - 999
Default value
200
Selection
mA | µA | nA | pA
Default value
pA
Sweep direction
Specification as to whether the substance is to be evaluated during for-
ward or reverse movement of the cyclical sweep.
Selection
anodic | cathodic
Default value
anodic
Navigation bar
Navigation bar for changing the displayed substances with the following
functions:
Jump to the first substance. Altered data is adopted.
Jump to the previous substance. Altered data is adopted.
Jump to the next substance (in the case of the last substance, jump to a
new, empty substance). Altered data is adopted.
Jump to the last substance. Altered data is adopted.