3 track, Track – Metrohm viva 1.0 Manual User Manual
Page 401
■■■■■■■■■■■■■■■■■■■■■■
5 Method
viva 1.0
■■■■■■■■
389
Action
NOTE
Is only displayed for sample data variables of the Number or Date/
Time type.
Action
on | off (Default value: off)
If this option is enabled, then one of the two following actions will auto-
matically be executed if the limit values are breached.
Selection
Cancel determination | Cancel determination
and series
Default value
Cancel determination
Cancel determination
If this option is selected, then the ongoing determination will be can-
celed (without database entry). The next determination of the series
will then be started. If an exit track is present, it will be executed. If an
error track is present, it will not be executed.
Cancel determination and series
If this option is selected, then the ongoing determination will be can-
celed (without database entry). The next determination of the series
will not be started. If an exit track is present, it will be executed. If an
error track is present, it will not be executed.
5.5.2.2.3
TRACK
Dialog window: Method
▶ TRACK ▶ Properties... ▶ TRACK - 'Command name'
Start command for normal track.
Appearance
The command has the following appearance:
Command name
Name of the command.
Entry
1 - 25 characters
Default value
Track %1