11 rack data subwindow, 1 configuration - rack data, 2 rack table – Metrohm viva 1.0 Manual User Manual
Page 731: 1 rack table, 11 rack data subwindow 6.11.1, Configuration - rack data, Rack table, Rack data
■■■■■■■■■■■■■■■■■■■■■■
6 Configuration
viva 1.0
■■■■■■■■
719
Upper limit
Upper warning limit for global variable.
Input range
-1.0E+99 - 1.0E+99 (max. 10 digits)
6.11
Rack data subwindow
6.11.1
Configuration - Rack data
Subwindow: Configuration
▶ Rack data
Rack data subwindow
The Rack data subwindow contains the rack table with all sample racks
for Metrohm sample changers defined for the client. It can be shown in
the Configuration program part as a part of the configuration view or (if
not present in the user interface) in a separate window with
View
▶ Quick access. The subwindow can be enlarged and reduced as
required; it can also be maximized.
Display of sample racks
The rack table is empty by default. Sample racks that are placed on sample
processors are recognized automatically and included in the rack table.
Additional racks can be imported manually from the program directory ...
\Metrohm\viva\res\rackdata.
6.11.2
Rack table
6.11.2.1
Rack table
Subwindow: Configuration
▶ Rack data
Contents
The rack data table shows the following information about the configured
racks as the default setting:
Rack name
Name of the rack.
Rack code
Rack code of the rack.
Number of positions
Number of sample positions on the rack.
Beaker radius samples
Radius of the beakers on the rack.