beautypg.com

2 call cond - properties – Metrohm viva 1.0 Manual User Manual

Page 431

background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

viva 1.0

■■■■■■■■

419

Identification

Description

FIN

Command status; 1 = Command has ended at
least once
, 0, invalid (variable not available) =
Command has never ended

LCO

Loop counter; current quantity of completed runs of
the called track

LST

Start time of the call command

SDV

Actual value of the standard deviation when the
command is canceled

5.5.2.4.6.2

CALL COND - Properties

Dialog window: Method

CALL COND Properties... CALL COND - 'Command

name'

Command name

Name of the command.

Entry

25 characters

Call text

Freely editable text for the designation of the call that is displayed in the
command depiction.

Entry

50 characters

Default value

'empty'

Track name

Selection of the VA track to be called up. All of the VA tracks available in
the method are displayed.

Selection

'Track name'

Stop criteria
Definition of four possible criteria for the cancellation of the command
call. The stop criterion that is reached first applies for the cancellation. If
not stop criterion is switched on, then the action will not be carried out.

Evaluation quantity

on | off (Default value: on)

If this check box is activated, then the call of the VA track will no longer
be carried out if the standard deviation of the evaluation quantity for the
selected substance falls below the specified value.