5 Method
viva 1.0
■■■■■■■■
419
Identification
Description
FIN
Command status; 1 = Command has ended at
least once, 0, invalid (variable not available) =
Command has never ended
LCO
Loop counter; current quantity of completed runs of
the called track
LST
Start time of the call command
SDV
Actual value of the standard deviation when the
command is canceled
5.5.2.4.6.2
CALL COND - Properties
Dialog window: Method
▶ CALL COND ▶ Properties... ▶ CALL COND - 'Command
name'
Command name
Name of the command.
Entry
25 characters
Call text
Freely editable text for the designation of the call that is displayed in the
command depiction.
Entry
50 characters
Default value
'empty'
Track name
Selection of the VA track to be called up. All of the VA tracks available in
the method are displayed.
Selection
'Track name'
Stop criteria
Definition of four possible criteria for the cancellation of the command
call. The stop criterion that is reached first applies for the cancellation. If
not stop criterion is switched on, then the action will not be carried out.
Evaluation quantity
on | off (Default value: on)
If this check box is activated, then the call of the VA track will no longer
be carried out if the standard deviation of the evaluation quantity for the
selected substance falls below the specified value.