3 meas ref - measuring parameters, Measuring parameters – Metrohm viva 1.0 Manual User Manual
Page 437
■■■■■■■■■■■■■■■■■■■■■■
5 Method
viva 1.0
■■■■■■■■
425
If the not defined option is selected in the Device name field, all instru-
ment types or groups with which the command can be executed can be
selected, regardless of the instruments present in the device table.
Selection
Avantes Spectrometer
Default value
Avantes Spectrometer
5.5.2.5.2.3
MEAS Ref - Measuring parameters
Tab: Method
▶ MEAS Ref ▶ Properties... ▶ MEAS Ref - 'Command name' ▶ Mea-
suring parameters
Command name
Name of the command.
Entry
25 characters
The parameters for the measuring procedure can be defined on this tab.
Start wavelength
Lower limit of the spectrum.
Input range
100.0 - 2,000.0 nm
Default value
400.0 nm
End wavelength
Upper limit of the spectrum.
Input range
100.0 - 2,000.0 nm
Default value
1,000.0 nm
Integration time
Integration time for the acquisition of the spectrum.
Input range
0.01 - 600,000 ms
Default value
6 ms
Averaged spectra
The number of spectra that are acquired and averaged.
Input range
1 - 10,000
Default value
10
Smoothing
Number of neighboring pixels with which the value for each measurement
pixel is determined.
Input range
0 - 100 pixels
Default value
0 pixels