5 properties - glp, Properties - glp – Metrohm viva 1.0 Manual User Manual
Page 743
■■■■■■■■■■■■■■■■■■■■■■
7 Devices
viva 1.0
■■■■■■■■
731
Acoustic signal
on | off (Default value: off)
If this check box is activated, an acoustic signal will be emitted in addition
to the message.
Action
The actions are only editable if Monitor calibrator is enabled.
If it is found during calibrator monitoring that the replacement interval has
expired, one of the following actions will be triggered automatically dur-
ing the start test:
Selection
Record message | Display message | Cancel
determination
Default value
Display message
Record message
The message that the replacement interval of the calibrator has expired
will be automatically saved with the determination.
Display message
A message is displayed and you can select whether you want to con-
tinue the run or cancel it. If the run is continued, the message that the
replacement interval of the calibrator has expired will be automatically
saved with the determination.
Cancel determination
The ongoing determination will be automatically canceled. The follow-
ing message must be confirmed with [OK].
7.1.5
Properties - GLP
Tab: Configuration
▶ Devices ▶ [Edit] ▶ Properties... ▶ Properties - 'Device type' -
'Device name'
▶ GLP
Tab: Configuration
▶ Dosing units ▶ [Edit] ▶ Properties... ▶ Dosing unit - 'Na-
me'
▶ GLP
Tab: Configuration
▶ Solutions ▶ [Edit] ▶ Properties... ▶ Solution - 'Name' ▶ GLP
GLP test date
Date of the last GLP test. This date can be selected by clicking on
in
the Select date dialog window (see Chapter 2.4.1, page 67).
Comment on GLP test
Comment regarding the GLP test.
Entry
1,000 characters