4 series start track, 5 series end track, Series start track – Metrohm viva 1.0 Manual User Manual
Page 386: Series end track
5.5 Method run subwindow
■■■■■■■■■■■■■■■■■■■■■■
374
■■■■■■■■
viva 1.0
5.5.1.2.4
Series start track
Subwindow: Method
A series start track is a track that can be created manually by the user
with Insert
▶ New track... ▶ Series start track or with the symbol.
It starts with the SERIES START command (see Chapter 5.5.2.2.5, page
391) and ends with the END command. Only one series start track can be
inserted in a given method.
The series start track is executed in the run only once at the beginning of
the first determination of a series and before the main track. This track can
be used to initialize, e.g. sample changers at the beginning of a series.
NOTE
No voltammetry commands (CVS) can be inserted in a series start track.
5.5.1.2.5
Series end track
Subwindow: Method
A series end track is a track that can be created manually by the user
with Insert
▶ New track... ▶ Series end track or with the symbol.
It starts with the SERIES END command (see Chapter 5.5.2.2.6, page
392) and ends with the END command. Only one series end track can be
inserted in a given method.
The series end track is executed in the run only once at the end of the last
determination of a series after the main track. This track can be used, e.g.
to set sample changers to the required state at the end of a series.