beautypg.com

Metrohm viva 1.0 Manual User Manual

Page 451

background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

viva 1.0

■■■■■■■■

439

[Delete]

Deletes the selected line.

Minimum evaluation

Minimum evaluation

on | off (Default value: off)

If this check box is activated, then the associated time for the minimum
measured value will be interpolated from the measuring point list.

Threshold value

The evaluation of the minimum begins as soon as the slope of the curve
exceeds the set threshold value.

Input range

0.1 - 1,000.0 mAU/s

Default value

20.0 mAU/s

Maximum evaluation

Maximum evaluation

on | off (Default value: off)

If this check box is activated, then the associated time for the maximum
measured value will be interpolated from the measuring point list.

Threshold value

The evaluation of the maximum begins as soon as the slope of the curve
exceeds the set threshold value.

Input range

0.1 - 1,000.0 mAU/s

Default value

20.0 mAU/s

Break point evaluation

Break point evaluation

on | off (Default value: off)

If this check box is activated, then sharp, almost 90° changes in directions
in the measurement curve will be evaluated.

EP criterion

Measure of the minimum sharpness of the break point. The smaller the EP
criterion set, the more break points will be found. As this is a relative value
related to the total measured value modification, even small changes in
the measured value can be evaluated as a break point for a small mea-
sured value range.