6 changing a remark, 7 single determination - determination run – Metrohm viva 1.0 Manual User Manual
Page 125
■■■■■■■■■■■■■■■■■■■■■■
3 Workplace
viva 1.0
■■■■■■■■
113
Analytical volume
Volume portion (aliquot) of the diluted sample's dilution volume used for
the determination (see Glossary, page 865). This volume is added to the
measuring vessel with the ADD SAMPLE command.
Input range
-1E+99 - 1E+99 (max. 10 places) mL
Default value
'empty' mL
Dilution volume
Maximum filling volume used for dilution of the sample (see Glossary,
page 868).
Input range
-1E+99 - 1E+99 (max. 10 places) mL
Default value
'empty' mL
NOTE
The name of the sample data variables assigned in the START com-
mand is automatically displayed for the sample data variables. Gener-
ally, no formula input is possible in these fields.
3.6.3.6
Changing a remark
Dialog window: Workplace
▶ Run ▶ Single determination ▶ Live modifica-
tions
▶ Modify remark
The Live modifications dialog window, in which the remark can be
modified live during an ongoing determination, opens with the context-
sensitive Modify remark menu item.
Remark
Freely definable remark on the determination, which is saved together
with the determination. Predetermined text templates may also be
selected instead of a manual entry (see Chapter 3.5.1, page 105).
Entry
100 characters
3.6.3.7
Single determination - Determination run
Tab: Workplace
▶ Run ▶ Single determination
A determination that is started on the Single determination tab in the
Run subwindow proceeds as follows:
1 - Loading the sample data
The sample data entered for the determination is loaded.
2 - Method selection
The following actions are triggered during method selection: