beautypg.com

5 dt calibration curve properties – Metrohm viva 1.0 Manual User Manual

Page 581

background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

viva 1.0

■■■■■■■■

569

5.6.5.5

DT calibration curve properties

Dialog window: Method

Evaluation - Calibration Calibration

curves

[Edit] Properties... Properties calibration

The calibration curve parameters for the substance selected in the table
can be edited in the Properties calibration dialog window.

Measuring command

Shows the selected measuring command.

Substance

Shows the selected substance.

Evaluation quantity

Selection of the size according to which the substance is to be evaluated.

Selection

Area | Height

Default value

Area

Weighting

on | off (Default value: on)

If this check box is activated, then the points on the calibration curve will
be weighted in accordance with the formula 1/evaluation quantity

².

Evaluation ratio

Evaluation ratio Q/Q(0) with which the calibration factor is calculated.

Input range

0.01 - 1.10

Default value

0.50

Evaluation parameters for standard

Curve type

Curve type for the evaluation of the calibration curve. Determines the type
according to which the calibration curve is to be fitted. A minimum num-
ber p of different effective addition volumes must exist for the standard
solution for the various curve types so that it is possible to fit this calibra-
tion curve. The minimum number p corresponds to the number of terms
in the calibration function.

Selection

Linear regression | Quadratic regression | Non-
linear regression
| Linear interpolation

Default value

Linear regression

Evaluation start

Evaluation ratio Q/Q(0) with which the calculation of the regression for
the standards is to be started.