beautypg.com

3 properties - calibration data, Calibration data – Metrohm viva 1.0 Manual User Manual

Page 705

background image

■■■■■■■■■■■■■■■■■■■■■■

6 Configuration

viva 1.0

■■■■■■■■

693

6.8.3.3

Properties - Calibration data

Tab: Configuration

Colorimetric sensors Edit Properties... Calibration

data

Wavelength

Wavelength at which the calibration of the instrument has been carried
out.

Input range

100.0 - 2,000.0 nm

Default value

400.0 nm

c0

Zero-order calibration coefficient.

Input range

-1E+99 - 1E+99 (max. 10 significant places)

Default value

0

c1

First-order calibration coefficient.

Input range

-1E+99 - 1E+99 (max. 10 significant places)

Default value

0

c2

Second-order calibration coefficient.

Input range

-1E+99 - 1E+99 (max. 10 significant places)

Default value

0

c3

Third-order calibration coefficient.

Input range

-1E+99 - 1E+99 (max. 10 significant places)

Default value

0