3 evaluation - substances, 1 substances - general, 2 substances - recognition – Metrohm viva 1.0 Manual User Manual
Page 567: Evaluation - substances, Substances

■■■■■■■■■■■■■■■■■■■■■■
5 Method
viva 1.0
■■■■■■■■
555
5.6.3
Evaluation - Substances
5.6.3.1
Substances - General
Subwindow section: Method
▶ Evaluation ▶ Substances
Application
The parameters of the substances that are to be determined are defined in
the Substances section. These are specific for each measuring command
with which the substances are to be determined. The measuring com-
mand is selected in the upper area of the dialog. Settings that have
already been made are retained when switching between different mea-
suring commands.
Selection of the measuring command
Measuring command
Selection list with all of the VA measuring commands with data acquisi-
tion defined in the method.
Selection
'VA measuring command'
Tabs
■
■
5.6.3.2
Substances - Recognition
Subwindow section: Method
▶ Evaluation ▶ Substances ▶ Recognition
Table
The table on the Recognition tab contains an overview of the parame-
ters that define the assignment of the substances to the peaks/steps that
are found. The table cannot be edited directly and contains the following
columns:
Substance
Name of the substance.
Active
Display as to whether the evaluation for this substance is active or not
active.
Characteristic potential
Verification potential for the assignment of a peak to a substance.