5 meas opt - additional measured values, Additional measured values – Metrohm viva 1.0 Manual User Manual
Page 453
■■■■■■■■■■■■■■■■■■■■■■
5 Method
viva 1.0
■■■■■■■■
441
Window = Time
Input range
0 - 999,999 s
Default value
999,999 s
5.5.2.5.4.5
MEAS Opt - Additional measured values
Tab: Method
▶ MEAS Opt ▶ Properties... ▶ MEAS Opt - 'Command name' ▶ Addi-
tional measured values
Command name
Name of the command.
Entry
25 characters
A maximum of 6 additional measured values can be defined on this tab.
These values can then be saved together in additional measured value col-
umns with the measured values present in the default settings.
Additional calculated measured values
Additional calculated measured values
on | off (Default value: off)
If this check box is activated, then a maximum of three new measured val-
ues can be calculated and saved from existing measured values that can
also be displayed as curves under the designation Calculated 1 - 3 and
be used as variables 'Command name.CA1 - 3' in formulas.
[New]
Opens the Calc. measured value # dialog window, in which the param-
eters for the additional measured value can be entered (see Chapter
5.5.2.5.7.1, page 457).
[Properties]
Opens the Calc. measured value # dialog window, in which the param-
eters for the additional measured value can be edited (see Chapter
5.5.2.5.7.1, page 457).
[Delete]
Deletes the calculated measured value selected in the table.
Additional external measured values
Additional external measured values
on | off (Default value: off)
If this check box is activated, then a maximum of three new measured val-
ues can be adopted from other measuring commands (e.g. MEAS) and
saved that can also be displayed as curves under the designation External
1 - 3 and be used as variables 'Command name.EX1 - 3' in formulas.