beautypg.com

5 meas opt - additional measured values, Additional measured values – Metrohm viva 1.0 Manual User Manual

Page 453

background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

viva 1.0

■■■■■■■■

441

Window = Time
Input range

0 - 999,999 s

Default value

999,999 s

5.5.2.5.4.5

MEAS Opt - Additional measured values

Tab: Method

MEAS Opt Properties... MEAS Opt - 'Command name' Addi-

tional measured values

Command name

Name of the command.

Entry

25 characters

A maximum of 6 additional measured values can be defined on this tab.
These values can then be saved together in additional measured value col-
umns with the measured values present in the default settings.

Additional calculated measured values

Additional calculated measured values

on | off (Default value: off)

If this check box is activated, then a maximum of three new measured val-
ues can be calculated and saved from existing measured values that can
also be displayed as curves under the designation Calculated 1 - 3 and
be used as variables 'Command name.CA1 - 3' in formulas.

[New]

Opens the Calc. measured value # dialog window, in which the param-
eters for the additional measured value can be entered (see Chapter
5.5.2.5.7.1, page 457)
.

[Properties]

Opens the Calc. measured value # dialog window, in which the param-
eters for the additional measured value can be edited (see Chapter
5.5.2.5.7.1, page 457)
.

[Delete]

Deletes the calculated measured value selected in the table.

Additional external measured values

Additional external measured values

on | off (Default value: off)

If this check box is activated, then a maximum of three new measured val-
ues can be adopted from other measuring commands (e.g. MEAS) and
saved that can also be displayed as curves under the designation External
1 - 3
and be used as variables 'Command name.EX1 - 3' in formulas.