Non-executed commands
The commands REQUEST, WAIT, RECEIVE, SEND, TRANSFER,
STIR, SCAN, CTRL, MOVE, SWING, LIFT, PUMP, RACK, HEATER,
ADD AUX, ADD STD, ADD SAMPLE, ADD SAMPLE DT, LQH,
PREP and EMPTY are run through but no longer performed.

Reports
The reports defined in the method are generated, but the defined out-
put targets are ignored. A report can only be correctly generated dur-
ing reprocessing if the defined report template has already been used
in the regular determination run. If the required report template is not
present in the original determination, a message entry is made, the
report is skipped and reprocessing is continued.

Reprocessing determinations with calibration method DT

– Determinations of the type Sample can be recalculated with

various calibrations. However, the determinations and the cali-
bration must have been carried out with the same method. The
method version may differ.

– In order for the current calibration data to be used, viva auto-

matically recalculates the determination of the type Standard
first after sample and standard selection.

– In order for determinations of the type Sample to be repro-

cessed with a new, recalculated calibration, the selection of
determinations may contain only one determination of the type
Standard.

– If determinations of the type Sample are to be recalculated,

new calibration data will only be used if the prompt is con-
firmed. If you click on [No], then the original calibration data will
be used.

– Determinations of the type Sample can be recalculated without

requiring a determination of the type Standard to be selected.
In this case, the original calibration data will be used.

4.5.2.5.4

Modifications

4.5.2.5.4.1

Modifications - Overview

Subwindow: Database

Determination overview Determinations Repro-

cess...

Reprocessing Modifications

Modifications can be made on the following tabs in the Modifications
subwindow of the Reprocessing dialog window:

Method
Modification of the method used in the selected determination.

Variables
Modification of the variables used in the selected determination.