100 m – EXFO CableSHARK P3 VF/DSL Cable Qualifier User Manual

Page 157

background image

CableSHARK P3 User Guide

145


The Isolation Resistance test setup options are described as follows:

TEST TIME (SEC):

The user can specify the “soak” time or the duration of the test in which 100 VDC
is applied to the circuit. Allowable times are from 1 to 99 seconds.

CONNECTION:

The user can select ALL, TIP-RING, TIP-GROUND, and RING-GROUND for
which to perform the test on.

8.9.2 Running an Isolation Resistance Test

Use caution when connecting the test set, as telephone cables may have dangerous A.C. or
D.C. voltages on them. The Isolation Resistance Test injects 100 VDC onto the line however
it is current limited.


Press the function button corresponding to the RUN / STOP button on the front of the CableSHARK to
initiate the test. Figure 8.9.2A shows an example of the results obtained from an Isolation Resistance Test.

(T-R) Resistance =
(T-G) Resistance =

(R-G) Resistance =

E

F

CHG

> 100 M
> 100 M
Not chosen

WARNING
Measuring Voltage: 100 VOLTS

!

Time Remaining =

00:00:25

MAIN
MENU

SAVE TO
MEM

SAVE TO

USB

UPLOAD
RESULT

Resistance

(T - R)

> 100 M

TIP-GND