Deactivate
Deactivating a monitor occurs when Profile sends a message to the 
monitor to discontinue measuring and logging flow data. ADS 
recommends deactivating a monitor upon completion of a project 
and/or when removing a monitor from service for an extended period 
of time. Deactivate a monitor in the following way: 
1. Select the Connect button to establish communication with the
monitor.
Profile initiates communication with the monitor and establishes a 
connection. 
2. Select Deactivate from the Functions drop-down list, and then
select the Perform button.
Profile deactivates the monitor.
Monitor Status
Check the monitor status to access the current system information 
regarding the available stored flow data, current sensor readings, 
communication status, hardware versions, data log rate, and battery 
voltage in the following way: 
Note: The content of the status report varies based on the 
monitor series and communication method. 
1. Select the Connect button to establish communication with the
monitor.
Profile initiates communication with the monitor and establishes a 
connection. Choose the Abort button prior to establishing a 
connection to abort the communication attempt. 
2. Select Monitor Status from the Functions drop-down list, and
then select the Perform button.
The Print Preview dialog displays the current system information in 
printable format. 