5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

958

■■■■■■■■

tiamo 2.2

Measurement with drift control
If the option Measurement with drift control has been selected, then
the following parameters will be displayed:

Measured value drift

The measured value is only accepted if the signal drift defined here has been
fallen short of.

Range

0.1 ... 999.0 mV/min

Default value

10.0 mV/min

Min. waiting time

The measured value is not accepted until the minimum waiting time has
elapsed, even if the signal drift has already been reached. The drift continues
to be checked while the waiting time is elapsing.

Range

0 ... 999999 s

Default value

0 s

Max. waiting time

If the signal drift has not yet been reached, then the measured value will be
accepted when the maximum waiting time has elapsed. If the waiting time
has not been newly entered then a waiting time that is suitable for the drift
will be calculated automatically according to the following equation:

Range

0 ... 999999 s

Default value

52 s

Measuring interval

Time interval for entering a measuring point in the measuring point list.

Range

0.1 ... 999999.0 s

Default value

2.0 s

Stop measured value

Stops when the preset measured value has been reached since the start of
the measurement.

Range

0 ... 1E+99

Selection

off

Default value

off

off
No stop.