2 properties curve evaluation - x axis, X axis – Metrohm tiamo 2.2 Manual User Manual
Page 329
■■■■■■■■■■■■■■■■■■■■■■
4 Database
tiamo 2.2
■■■■■■■■
315
4.5.2.6.4.6.4.2
Properties curve evaluation - x axis
Tab: Database
▶ Determinations ▶ Reprocess... ▶ Reprocessing ▶ Curve evalua-
tion
▶ [Edit] ▶ Curve evaluation ▶ Properties ▶ x axis
On the tab x axis the parameters for the graphical display of the curves on
the x axis can be set.
Command type
Shows the type of command for which the curve properties can be defined.
The curve properties defined for each command type for the reprocessing
window are saved per client.
Autoscaling
on | off (Default value: on)
If this option is switched on then all axes in the curve window will be scaled
automatically. In this case the fields Start value and End value cannot be
edited.
Size
Selection
Command-dependent selection
Command-dependent selection
Selection of the quantity to be shown on the x axis.
Label
Freely definable axis label for the x axis.
Selection
50 characters | auto
Default value
auto
auto
The designation from the field Size will be used.
Scaling
Fixed scaling of the x axis between the start value and the end value.
Start value
Initial value for scaling the x axis.
Only editable when the option Autoscaling is disabled.
Range
-1.0E12 ... 1.0E12
Default value
-1.0E12
End value
End value for scaling the x axis.