5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1080
■■■■■■■■
tiamo 2.2
Dosing rate
Rate at which dosing should take place. The maximum dosing rate depends
on the cylinder volume of the exchange unit or dosing device used. If the
entered dosing rate is too high for the selected dosing device, it will auto-
matically be reduced during dosing to the largest possible value. This
parameter is only displayed for Dosing criterion = Volume/Dosing
rate or Dosing rate/Dosing time.
Titrando, 855
Range
0.01 ... 166.00 mL/min
Default value
maximum mL/min
Titrino
Range
0.01 ... 150.00 mL/min
Default value
maximum mL/min
Note
The dosing rate should be reduced for viscous liquids.
Dosing time
Time during which dosing is to be carried out. This parameter is only dis-
played for Dosing criterion = Volume/Dosing time or Dosing rate/
Dosing time.
Range
0 ... 9999999 s
Default value
100 s
Pause
Waiting period before dosing is started.
Range
0 ... 999999 s
Default value
0 s
Temperature
Measuring temperature which can be entered manually. If a temperature
sensor is connected and the Temperature measurement is set on the
General/Hardware tab under Sensor to automatic or continuous, then
the temperature will be measured continuously. This value is used for tem-
perature correction during monitoring with pH measurements.
Titrando, 855
Range
-20.0 ... 150.0 °C
Default value
25.0 °C