5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
926
■■■■■■■■
tiamo 2.2
U(pol)
The polarization voltage is the voltage applied to the polarized electrode
during an amperometric measurement.
Titrando, 855, 867
Range
-1250 ... 1250 mV (Increment: 25)
Default value
400 mV
Titrino
Range
-1270 ... 1270 mV (Increment: 10)
Default value
400 mV
Electrode test
on | off (Default value: off)
If this option is enabled, then an electrode test will be carried out for polar-
izable electrodes during the transition from an inactive condition to a
measurement. A check is also made while doing so to ensure that the elec-
trode is properly connected and that no short-circuit is present.
Temperature measurement
Type of temperature measurement.
Titrando, 855, 867
Selection
continuous | automatic | off
Default value
automatic
continuous
A temperature sensor must be connected. The temperature will then be
measured continuously.
automatic
If a temperature sensor is connected then the temperature will be meas-
ured continuously. Otherwise, the temperature entered manually under
Temperature on the Measuring parameters tab will be used.
off
The temperature will not be measured. The temperature manually
entered under Temperature on the Measuring parameters tab will
be used.
Stirrer
Stirrer
Selection of the stirrer.
Titrando, 855, 867
Selection
1 | 2 | 3 | 4 | off