2 properties - global variable – Metrohm tiamo 2.2 Manual User Manual
Page 1339
■■■■■■■■■■■■■■■■■■■■■■
6 Configuration
tiamo 2.2
■■■■■■■■
1325
6.9.3.2
Properties - Global variable
Tab: Configuration
▶ Global variables ▶ [Edit] ▶ Properties... ▶ Global variable -
'Name'
▶ Global variable
Name
Name of global variable.
Input
50 characters
Type
Selection of the type for a new global variable. For existing global variables
the type will only be shown; it cannot be edited.
Selection
Number | Text | Date/Time
Default value
Number
Value
Value of the global variable. This value can be assigned in a method by a
CALC command or entered manually. For methods that use the global var-
iable it is available as the variable 'GV.Name.VAL' or 'GV.Name' (short
form) for calculations.
For type Date/Time variables, the date can be selected by pressing
in
the dialog window Select date.
Type = Number
Range
-1E+99 ... 1E+99 (max. 15 places)
Type = Text
Input
256 characters
Type = Date/Time
Selection
Date selection
Value (unit)
Designation of the unit. If a value is assigned automatically then the unit will
be automatically entered here. It is available for methods that use the global
variable as variable 'GV.Name.UNI' for calculations.
Input
20 characters
Selection
'empty' | mol/L | mmol/L | µmol/mL | g/L | mg/L
| µg/L | mg/mL | ppm | % | mEq/L | mL | µg
Default value
mol/L
Comment
It is possible to enter remarks about the global variable.