Metrohm tiamo 2.2 Manual User Manual
Page 958
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
944
■■■■■■■■
tiamo 2.2
Minimum evaluation
Minimum evaluation
on | off (Default value: off)
If this option is enabled, then the associated time for the minimum measured
value will be interpolated from the measuring point list.
Threshold value
The evaluation of the minimum begins as soon as the slope of the curve
exceeds the set threshold value.
Range
0.1 ... 20.0 °C/s
Default value
1.0 °C/s
Maximum evaluation
Maximum evaluation
on | off (Default value: off)
If this option is enabled, then the associated time and temperature for the
maximum measured value will be interpolated from the measuring point list.
Threshold value
The evaluation of the maximum begins as soon as the slope of the curve
exceeds the set threshold value.
Range
0.1 ... 20.0 °C/s
Default value
1.0 °C/s
Break-point evaluation
Note
This evaluation method is possible only with Titrando, 855, 856 and 867.
Break-point evaluation
on | off (Default value: off)
If this option is enabled, then sharp, almost 90° changes in directions in the
measurement curve will be evaluated.
EP criterion
Measure of the minimum sharpness of the break point. The smaller the EP
criterion set, the more break-points will be found. As this is a relative value
related to the total measured value modification, even small changes in the