Page 479: Det u

background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.2

■■■■■■■■

465

last
Only the last endpoint to be found will be recognized.
ascending
Only endpoints with a positive slope of the titration curve will be recog-
nized.
descending
Only endpoints with a negative slope of the titration curve will be rec-
ognized.

5.6.3.2.2.11

DET pH - Fix endpoint evaluation

Dialog window: Method

DET pH Properties... Additional evaluations Fix

endpoint evaluation #

Quantity

Selection of the fixed measured quantity to which the associated values for
the other quantities are to be interpolated from the list of measured points.

Selection

Measured value | Time | Volume

Default value

Measured value

Fix value

Value of the fix endpoint.

Measured value
Range

-20.000 ... 20.000 pH

Time
Range

0.0 ... 999999.9 s

Volume
Range

0.00000 ... 9999.99 mL

5.6.3.2.3

DET U

5.6.3.2.3.1

DET U - Overview

Dialog window: Method

DET U Properties... DET U - 'Command name'

Command for dynamic equivalence point titrations with potentiomet-
ric voltage measurement.

Devices
This command can be executed with the following devices:

Titrando: 808, 809, 835, 836, 857, 888, 904, 905, 906, 907

Titrino: 716, 721, 736, 751, 785, 794, 798, 799

Robotic Titrosampler: 855