5 Method
tiamo 2.2
■■■■■■■■
465
last
Only the last endpoint to be found will be recognized.
ascending
Only endpoints with a positive slope of the titration curve will be recog-
nized.
descending
Only endpoints with a negative slope of the titration curve will be rec-
ognized.
5.6.3.2.2.11
DET pH - Fix endpoint evaluation
Dialog window: Method
▶ DET pH ▶ Properties... ▶ Additional evaluations ▶ Fix
endpoint evaluation #
Quantity
Selection of the fixed measured quantity to which the associated values for
the other quantities are to be interpolated from the list of measured points.
Selection
Measured value | Time | Volume
Default value
Measured value
Fix value
Value of the fix endpoint.
Measured value
Range
-20.000 ... 20.000 pH
Time
Range
0.0 ... 999999.9 s
Volume
Range
0.00000 ... 9999.99 mL
5.6.3.2.3
DET U
5.6.3.2.3.1
DET U - Overview
Dialog window: Method
▶ DET U ▶ Properties... ▶ DET U - 'Command name'
Command for dynamic equivalence point titrations with potentiomet-
ric voltage measurement.
Devices
This command can be executed with the following devices:
Titrando: 808, 809, 835, 836, 857, 888, 904, 905, 906, 907
Titrino: 716, 721, 736, 751, 785, 794, 798, 799
Robotic Titrosampler: 855