5 kft upol - titration parameters, Titration parameters – Metrohm tiamo 2.2 Manual User Manual
Page 783
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.2
■■■■■■■■
769
Stop time
The titration is stopped if the absolute stop time defined here has elapsed
since the start of the titration. This parameter is only shown for Stop cri-
terion = time and Delay time = inf.
Titrino
Range
1 ... 999999 s
5.6.3.5.3.5
KFT Upol - Titration parameters
Tab: Method
▶ KFT Upol ▶ Properties... ▶ Titration parameters
Command name
Name of the command.
Input
25 characters
The parameters for the running of the titration are defined on this tab.
Titration direction
Selection of the titration direction.
Selection
+ | - | auto
Default value
auto
+
positive measured value alteration
-
negative measured value alteration
auto
The titration direction is determined automatically from the start meas-
ured value and the set endpoint.
Extraction time
The titration is not stopped until the extraction time has elapsed (even when
the EP has already been reached). The entry of an extraction time may be
advisable, e.g. for the titration of sparingly soluble samples.
Range
0 ... 999999 s
Temperature
Manually entered titration temperature. If a temperature sensor is connected
and the Temperature measurement is set on the General/Hardware
tab under Sensor to automatic or continuous, then the temperature will
be measured continuously. This value is used for temperature correction in
pH measurements.