7.8 730 Sample Changer
■■■■■■■■■■■■■■■■■■■■■■
1384
■■■■■■■■
tiamo 2.2
Input
1000 characters
Default value
empty
7.8.3
Properties - Tower
Tab: Configuration
▶ Devices ▶ [Edit] ▶ Properties... ▶ Properties - 'Device type' -
'Device name'
▶ Towers
Note
This tab is only displayed if a PC and a sample changer are connected via
an RS-232 interface.
On the tab Towers the tower parameters for Tower 1 and Tower 2 (if pres-
ent) are shown for the 730 Sample Changer. These cannot be edited, but
must be set on the device itself via the keyboard.
Tower 1 / Tower 2
Max. stroke path
Shows the lowest permitted lift position for the two towers. A lift height of
0 mm corresponds to the "Home position", i.e. the lift is moved right to the
top.
Beaker sensor
Shows whether the beaker sensor for the two towers is switched on or off.
Each time a sample position is moved to the beaker sensor checks whether
a beaker is present or not.
Tower 1
Number of pumps
Shows how many pumps are connected to Tower 1.
759 Swing Head
Shows whether a swing head is mounted on Tower 1.
Tower 2
Number of pumps
Shows how many pumps are connected to Tower 1.
Note
A swing head can only be mounted on Tower 1.