6 met upol - potentiometric evaluation, Potentiometric evaluation – Metrohm tiamo 2.2 Manual User Manual
Page 652
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
638
■■■■■■■■
tiamo 2.2
Titrando, 855
Range
0.01000 ... 9999.99 mL
Selection
off
Default value
off
Stop time
Stops when the preset time (including start conditions) has elapsed since the
start of the titration.
Titrando, 855
Range
0 ... 999999 s
Selection
off
Default value
off
Filling rate
Speed with which the dosing cylinder is to be refilled after the titration. The
maximum filling rate depends on the cylinder volume of the exchange unit
or dosing device used.
Titrando, 855
Range
0.01 ... 166.00 mL/min
Selection
maximum
Default value
maximum
Titrino
Range
0.01 ... 150.00 mL/min
Selection
maximum
Default value
maximum
5.6.3.3.5.6
MET Upol - Potentiometric evaluation
Tab: Method
▶ MET Upol ▶ Properties... ▶ Potentiometric evaluation
Command name
Name of the command.
Input
25 characters
Parameters for the potentiometric evaluation of the titration curve with
automatic equivalence point recognition are defined on this tab. One of the
following three evaluation methods can be chosen here:
Potentiometric evaluation
Selection
Evaluation without window | Evaluation with
measured value window (Upol) | Evaluation
with volume window (mL)
Default value
Evaluation without window