Metrohm tiamo 2.2 Manual User Manual
Page 439
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.2
■■■■■■■■
425
Statistics
on | off (Default value: off)
If this parameter is switched on, then the statistic functions defined in the
CALC command will be calculated.
Note
In order to make sure that the statistics will really be calculated, the feature
must also be enabled in the Workplace program part on the Single
determination tab.
Number of single determinations
The number of single determinations defined here is valid for all CALC com-
mands in the method.
Range
1 ... 99
Selection
not defined
Default value
not defined
not defined
With not defined the number of single determinations must be specified
on the tab Determination series in the subwindow Run of the pro-
gram part Workplace.
Conditioning
Automatic conditioning
on | off (Default value: on)
If this option is enabled, the conditioning will be started automatically at
the start of a determination for all commands with activated conditioning
(SET, KFT, KFC and BRC). After the track with the conditioning command
has been finished, conditioning is started again automatically and will be
continued even after the end of the determination.
If this parameter is disabled, then conditioning will not be started until the
start of the command. Conditioning is not started again after the end of the
track with the conditioning command. This parameter must be switched off
for automation applications.