5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
484
■■■■■■■■
tiamo 2.2
The following additional methods for evaluation of titration curves can be
activated and defined on this tab:
Fix endpoint evaluation
Fix endpoint evaluation
on | off (Default value: off)
If this check box is activated, then the associated values for the other meas-
ured quantities will be interpolated with a fixed measured quantity (Meas-
ured value, Volume or Time) for the fix endpoint from the measuring
point list.
Moves the selected line up (changes the sequence).
Moves the selected line down (changes the sequence).
[New]
Opens the dialog window Fix endpoint evaluation # to enter a new fixed
endpoint (see Chapter 5.6.3.2.3.11, page 491).
[Properties]
Opens the dialog window Fix endpoint evaluation # to edit the fixed
endpoint selected in the table (see Chapter 5.6.3.2.3.11, page 491).
[Delete]
Deletes the selected line.
pK/HNP evaluation
pK/HNP evaluation
on | off (Default value: off)
If this option is enabled, then the pK value is determined from the titration
curve which corresponds to the pH value at the half neutralization point
(see Chapter 5.6.3.10.1, page 875).
Minimum evaluation
Minimum evaluation
on | off (Default value: off)
If this option is enabled, then the associated volume, time and temperature
for the minimum measured value will be interpolated from the measuring
point list.