6 Configuration

tiamo 2.2

■■■■■■■■

1299

Note

The days are monitored, e.g. the calibration becomes invalid when the
date changes and not at the time time of day that calibration took place.

Can only be edited if monitoring is switched on.
Range

0 ... 999 days

Default value

999 days

Next calibration

Date on which the next calibration is to be carried out. The date can be

selected by pressing

in the dialog window Select date (see Chapter

2.5.1, page 83). When a date has been entered, the value for the Calibra-
tion interval
will be adapted automatically.

Can only be edited if monitoring is switched on.
Set value

      Calibration date + 999 days

Message
The message options can only be edited if monitoring is switched on.

Message by E-mail

on | off (Default value: off)

The message is additionally sent to the E-mail address defined under [E-
mail]
if this option is activated. The message is sent in text format.

[E-mail...]

[E-mail...] opens the window Send E-mail (see Chapter 2.6.1, page 86).

Acoustic signal

on | off (Default value: off)

If this option is enabled, an acoustic signal will be emitted additionally to the
message defined above

Action
The actions can only be edited if monitoring is switched on.

If during calibration monitoring it is found that its validity has expired then
one of the following actions will be triggered automatically at Start test:

Selection

Document message | Display message | Cancel
determination

Default value

Display message