Metrohm tiamo 2.2 Manual User Manual
Page 1482
8.4 Manual Control - Functions
■■■■■■■■■■■■■■■■■■■■■■
1468
■■■■■■■■
tiamo 2.2
Output signal
Current status
Displays the current status of the 14 output lines.
Templates/Input
Input of the binary pattern for the output signal or selecting a predefined
signal pattern.
It is possible to enter the following characters:
0 = line inactive
1 = line active
* = any line status
p = set pulse (pulse length = 200 ms. If a pulse of another length should be
put out, a corresponding template has to be defined.)
Titrando, 831, 846, 855, 856, 867, 751, 756, 758, 784, 785, 795, 798, 799,
sample changer
Selection
Bit-pattern with exactly 14 characters (0, 1, *,
p) | ************** | Signal pattern
Default value
**************
702, 716, 718, 719, 720, 721, 736, 794
Selection
Bit-pattern with exactly 8 characters (0, 1, *,
p) | ******** | Signal pattern
Default value
********
Spectrometer
Selection
Bit-pattern with exactly 10 characters (0, 1, *,
p) | ********** | Signal pattern
Default value
**********
The output lines are numbered from right to left:
Output 13 12 11 10 9 8 7 6 5 4 3 2 1 0
Bit
13 12 11 10 9 8 7 6 5 4 3 2 1 0
Examples:
************1* sets the output line 1 to enabled (= set), which would result
in a stop command for a connected Titrino, for example.
************0* sets the line to inactive.