7 det ph - additional evaluations, Additional evaluations – Metrohm tiamo 2.2 Manual User Manual
Page 471
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.2
■■■■■■■■
457
[New]
Opens the dialog window Measured value # in which the parameters for
a new window can be entered (see Chapter 5.6.3.2.2.9, page 462).
[Properties]
Opens the dialog window Measured value # in which the parameters for
the selected window can be edited (see Chapter 5.6.3.2.2.9, page 462).
[Delete]
Deletes the window selected in the table.
Evaluation with volume window (mL)
[New]
Opens the dialog window Volume window # in which the parameters for
a new window can be entered.
[Properties]
Opens the dialog window in which the parameters for the selected window
can be edited (see Chapter 5.6.3.2.2.10, page 464).
[Delete]
Deletes the window selected in the table.
5.6.3.2.2.7
DET pH - Additional evaluations
Tab: Method
▶ DET pH ▶ Properties... ▶ Additional evaluations
Command name
Name of the command.
Input
25 characters
The following additional methods for evaluation of titration curves can be
activated and defined on this tab:
Fix endpoint evaluation
Fix endpoint evaluation
on | off (Default value: off)
If this check box is activated, then the associated values for the other meas-
ured quantities will be interpolated with a fixed measured quantity (Meas-
ured value, Volume or Time) for the fix endpoint from the measuring
point list.
Moves the selected line up (changes the sequence).