2 cal cond - general/hardware – Metrohm tiamo 2.2 Manual User Manual
Page 1064
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1050
■■■■■■■■
tiamo 2.2
Identifica-
tion
Description
.LP.TEM
Temperature for the last measuring point of the measur-
ing point list in °C
.LP.TIM
Time in s until the last measuring point in the measuring
point list is reached
.MTE
Temperature measurement with sensor; 1 = on, 0 =
off)
.NMP
Number of measuring points in the measuring point list
.RTE
Reference temperature in °C (only MEAS Cond and CAL
Cond)
.STY
Type of stop with which the command was stopped: 1 =
stopped normally, 0 = stopped manually with
[Quit] or [Stop].
.TC
Temperature coefficient in %/°C
5.6.5.8.2
CAL Cond - General/Hardware
Tab: Method
▶ CAL Cond ▶ Properties... ▶ General/Hardware
Command name
Name of the command.
Input
25 characters
The general parameters for the device, the sensor and the stirrer are defined
on this tab.
Device
Device name
Selection of a device from those available in the device table. The only devi-
ces which shall be offered are those with which the command can run.
Selection
'Device name' | not defined
Default value
not defined
not defined
In this case, the device name must be assigned by the user at the start of
the method.
Device type
Display or selection of the device type.
Selection
712 Conductometer | 856 Conductivity Module
Default value
856 Conductivity Module