beautypg.com

Metrohm tiamo 2.2 Manual User Manual

Page 52

background image

2.4 Formula editor

■■■■■■■■■■■■■■■■■■■■■■

38

■■■■■■■■

tiamo 2.2

Identifica-
tion

Description

Commands

.HP{x}.MEA

Measured value for the HNP x (1…
9) in mV (HNP = half neutralization
potential)

DET, MET

.HP{x}.TEM

Temperature for the HNP x (1…9)
in °C

DET, MET

.HP{x}.TIM

Time in s until the HNP x (1...9) is
reached

DET, MET

.HP{x}.VOL

Volume for the HNP x (1…9) in mL

DET, MET

.IGF

Initial gas flow (measured value at
the time of the start of the com-
mand) in mL/min

MEAS T/Flow

.IME

Initial measured value (measured
value before start conditions are
processed) in the unit of the meas-
ured value

DET, MET, SET,
KFT, KFC, BRC,
STAT, MEAS
(without Ref,
Spec), CAL
MEAS, STDADD,
DOS

.ITE

Initial temperature (temperature
before start conditions are pro-
cessed) in °C

DET, MET, SET,
KFT, KFC, BRC,
STAT, MEAS
(without T/Flow,
Ref, Spec), CAL
MEAS, STDADD,
DOS

.LCO

Loop counter = current quantity of
completed runs, both of Repeat
loops
and of While loops

LOOP, CAL LOOP

.LP.CAx

Calculated value x (1…3) for the last
measuring point on the measuring
point list

DET, MET, SET,
KFT, KFC, BRC,
STAT, MEAS
(without Ref,
Spec, Opt), DOS

.LP.CHA

Load for the last measuring point of
the measuring point list in mA s

KFC, BRC

.LP.CNT

Intensity of the last measuring point
on the measuring point list in
counts

MEAS Opt