5 Method

tiamo 2.2

■■■■■■■■

897

Appearance
The command has the following appearance:

Parameters
The parameters for the command MEAS U are configured on the following
4 tabs:

General/Hardware
Parameters for devices, sensors and stirrers.

Measuring parameters
Parameters for setting the measurement procedure.

Evaluations
Defining further methods for evaluating measurement curves.

Additional measured values
Defining further measured values which can be saved as additional col-
umns in the measuring point list.

Command variables
The following command variables are generated in the method run by the
command MEAS U and can be used in formulas under the designation
'Command name.Variable designation':

Identification

Description

.BLV

Blank value of the sensor used for the command (only
for ISE sensors)

.BP{x}.MEA

Measured value for the break point x (1...9) in the unit
of the measured value

.BP{x}.TEM

Temperature for the break point x (1…9) in °C

.BP{x}.TIM

Time for the break point x (1…9) in s

.BSY

Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY
; invalid (variable not available) = Command
has never been started

.DBL

Total duration for the processing of the command in s

.EME

End measured value (measured value after processing
of the command) in the unit of the measured value

.ENP

Electrode slope of the sensor used for the command (in
mV for ISE sensors)