5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
868
■■■■■■■■
tiamo 2.2
Range
0.00000 ... 50.0000 mL
Default value
0 mL
Dosing rate
Rate (volume/time unit) at which the start volume is to be added. The max-
imum dosing rate depends on the cylinder volume of the dosing unit used.
Range
0.100 ... 166.000 mL/min
Selection
maximum
Default value
maximum
Pause
Pause
Waiting time, e.g. for the electrode to settle down after the start or a reac-
tion time after the dosing of a start volume. The pause follows at the end
of all the start conditions.
Range
0 ... 9999 s
Default value
0 s
5.6.3.9.4
TET - Titration parameters
Tab: Method
▶ TET ▶ Properties... ▶ Titration parameter
Command name
Name of the command.
Input
25 characters
The parameters for the running of the titration are defined on this tab.
Dosing
Dosing rate
Rate (volume/time unit) at which dosing is carried out. The maximum dosing
rate depends on the cylinder volume of the dosing unit used.
Range
0.100 ... 166,000 mL/min
Default value
2.000 mL/min
Selection
maximum
Filter
Filter factor
Factor which determines the digital filtering of the measuring curve. The
higher this value, the more the noise is damped. The filtering has an influence
on the shape of the peak and the EP shift.