4 properties - titer history, Titer history – Metrohm tiamo 2.2 Manual User Manual
Page 1288
6.6 Subwindow Titrants/Solutions
■■■■■■■■■■■■■■■■■■■■■■
1274
■■■■■■■■
tiamo 2.2
Selection
Record message | Display message | Stop deter-
mination
Default value
Display message
Record message
The message that the validity period has expired will be automatically
saved with the determination.
Display message
A message is displayed and you can choose whether you want to con-
tinue the run or cancel it. If the run is continued then the message that
the validity period has expired will be automatically saved with the deter-
mination.
Stop determination
The running determination will be automatically canceled. The following
message must be confirmed with [OK].
6.6.3.4
Properties - Titer history
Tab: Configuration
▶ Titrants/solutions ▶ Properties ▶ Titer history
History table
The table contains the last 10 titer determinations for the selected solution
and can neither be edited nor sorted. The titer determinations are arranged
chronologically so that the most recent determination is shown last.
Titer
Titer values are shown in the following colors:
■
Blue, if the titer has been determined automatically by a method.
■
Black, if the titer has been entered manually.
■
Orange, if the titer is outside the warning limits.
■
Red, if the titer is outside the intervention limits.
If the warning or intervention limits are exceeded then the line number will
also be shown with an orange and red background respectively.
Date titer det.
Date and time of the titer determination.
Titer method
Name of the method with which the titer determination was carried out. If
the titer has been entered manually then manual will be shown here.