5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
932
■■■■■■■■
tiamo 2.2
Maximum evaluation
Maximum evaluation
on | off (Default value: off)
If this option is enabled, then the associated time and temperature for the
maximum measured value will be interpolated from the measuring point list.
Threshold value
The evaluation of the maximum begins as soon as the slope of the curve
exceeds the set threshold value.
Range
0.5 ... 10.0 µA/s
Default value
5.0 µA/s
Break-point evaluation
Note
This evaluation method is possible only with Titrando, 855 and 867.
Break-point evaluation
on | off (Default value: off)
If this option is enabled, then sharp, almost 90° changes in directions in the
measurement curve will be evaluated.
EP criterion
Measure of the minimum sharpness of the break point. The smaller the EP
criterion set, the more break-points will be found. As this is a relative value
related to the total measured value modification, even small changes in the
measured value can be evaluated as a break point for a small measured value
range.
Range
0 ... 1.0
Default value
0.3
Slope
Minimum difference between the slope before and after the break point.
The smaller the difference, the more break-points will be found.
Range
0.0 ... 10.0
Default value
0.9
Smoothing factor