5 Method
tiamo 2.2
■■■■■■■■
753
Note
Conditioning must already be completed (= Condok) at the time the
SEND command is received, otherwise the command to start the titra-
tion will be ignored.
Show measured value during conditioning
on | off (Default value: off)
If this option is enabled, then the measured value will be shown in the sub-
window Live display of the program part Workplace, in addition to the
drift and the time.
5.6.3.5.2.8
KFT Ipol - Additional evaluations
Tab: Method
▶ KFT Ipol ▶ Properties... ▶ Additional evaluations
Command name
Name of the command.
Input
25 characters
The following additional methods for the evaluation of the titration curves
can be activated and defined on these tabs:
Fix endpoint evaluation
Fix endpoint evaluation
on | off (Default value: off)
If this check box is activated, then the associated values for the other meas-
ured quantities will be interpolated with a fixed measured quantity (Meas-
ured value, Volume or Time) for the fix endpoint from the measuring
point list.
Moves the selected line up (changes the sequence).
Moves the selected line down (changes the sequence).
[New]
Opens the dialog window Fix endpoint evaluation # to enter a new fixed
endpoint (see Chapter 5.6.3.5.2.10, page 756).
[Properties]
Opens the dialog window Fix endpoint evaluation # to edit the fixed
endpoint selected in the table (see Chapter 5.6.3.5.2.10, page 756).