3 kft upol - start conditions, Start conditions – Metrohm tiamo 2.2 Manual User Manual
Page 777
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.2
■■■■■■■■
763
Titrando
Selection
1 | 2 | 3 | 4 | off
off
No stirrer will be used.
Stirring rate
Setting the stirring rate. The algebraic sign of the stirring rate changes the
direction in which the stirring is done.
Titrando
Range
-15 ... 15
Default value
8
Switch off automatically
on | off (Default value: on)
If this option is enabled, the stirrer will be switched off automatically when
the command has finished. This parameter is displayed only for Titrandos.
Switch on/off automatically
on | off (Default value: on)
If this option is enabled, the stirrer will be switched on automatically at the
start of the command and will be switched off automatically at the end of
the command. This parameter is displayed only for 751, 785, 784, 795 and
799.
5.6.3.5.3.3
KFT Upol - Start conditions
Tab: Method
▶ KFT Upol ▶ Properties... ▶ Start conditions
Command name
Name of the command.
Input
25 characters
The start conditions are processed in the listed sequence before the titration
is started.
Initial measured value
Note
Is displayed only for Titrando.
The initial measured value is determined before the other start conditions
are processed. The measured value acceptance depends on the three fol-
lowing parameters: