5 result variables, Result vari- ables – Metrohm tiamo 2.2 Manual User Manual
Page 59
■■■■■■■■■■■■■■■■■■■■■■
2 General program functions
tiamo 2.2
■■■■■■■■
45
Identifica-
tion
Description
Commands
.TITER
Titer value of the solution used for
the command
DET, MET, SET,
KFT, STAT,
STDADD dos,
STDADD auto,
ADD, DOS, LQH
.TOL
Tolerance within which a measured
peak can be assigned to a calibra-
tion wavelength.
CAL Spec
.TOU
Timeout status: 1 = Timeout
expired; 0 = Timeout not
expired
RECEIVE, TRANS-
FER, SCAN
.UNI
Mapping unit which is defined in
the configuration for an IO port
(Text)
ANALOG IN,
ANALOG OUT
.VAL
Value of the command variables
(optional, i.e. Command-
name.Portname' =Command-
name.Portname.VAL') (Num-
ber). The value returned is the value
after the ending of the command.
ANALOG IN,
ANALOG OUT,
DIGITAL IN, DIG-
ITAL OUT
.VAR
Variance of the calculated result of
the standard addition in the selec-
ted unit
STDADD
.VOL
Dosed volume
STDADD, ADD,
DOS, LQH
2.4.3.5
Result variables
Dialog window: Formula Editor
▶ Variables
Result variables are method-specific and are defined in the CALC command
of the method. The field Variables of the formula editor lists all Result
variables which are available for the current method.
Syntax
'RS.Result name.Variable identification'
Examples: 'RS.RS01.VAL' (='RS.RS01'), 'RS.testTime.UNI'
You can select the method variables directly in the formula editor under
Variables
▶ Result Variables in order to avoid syntax errors.