2 add - general/hardware – Metrohm tiamo 2.2 Manual User Manual
Page 1076
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1062
■■■■■■■■
tiamo 2.2
Identifica-
tion
Description
.BSY
Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY; invalid (variable not available) = Command has
not ever been started
.CONC
Concentration of the solution used for the command
.CYL
Cylinder volume of the exchange or dosing unit used for
the command
.FIN
Command status; 1 = Command has been ended at
least once, 0, invalid (variable not available) = Com-
mand has never been ended
.TITER
Titer value of the solution used for the command
.VOL
Currently dosed volume
5.6.6.2.2
ADD - General/Hardware
Tab: Method
▶ ADD ▶ Properties... ▶ General/Hardware
Command name
Name of the command.
Input
25 characters
The general parameters for the device, the dosing device and the stirrer are
defined on this tab.
Device
Device name
Selection of a device from those available in the device table. The only devi-
ces which shall be offered are those with which the command can run.
Selection
'Device name' | not defined
Default value
not defined
not defined
In this case, the device name must be assigned by the user at the start of
the method.
Device type
Display or selection of the device type. If a device is selected under Device
name, then this Device type field can no longer be edited, and the device
type belonging to the selected device is displayed instead
If the option not defined is selected as Device name, any device types or
device groups which are able to execute the command can be selected,
independently of the devices in the device table.