Metrohm tiamo 2.2 Manual User Manual
Page 1071
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.2
■■■■■■■■
1057
Appearance
The command has the following appearance:
Parameters
The parameters for the command CAL Cond are configured on the follow-
ing 3 tabs:
■
General/Hardware
Parameters for device.
■
Measuring parameters
Parameters for setting the measurement procedure.
■
Calibration wavelengths
Input of the reference wavelengths. Definition of the behavior of the
calibration.
Command variables
The following command variables are generated in the method run by the
command CAL Spec and can be used in formulas under the designation
'Command name.Variable designation':
Identification
Description
.BSY
Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY; invalid (variable not available) = Command
has never been started
.C0
Coefficient c0 of the calibration curve
.C1
Coefficient c1 of the calibration curve
.C2
Coefficient c2 of the calibration curve
.C3
Coefficient c3 of the calibration curve
.CAL
Status of the calibration: 1 = calibrated normally, 0
= canceled, because reference wavelength was not
found.
.DBL
Total duration for the processing of the command in s
.STY
Type of stop with which the command was stopped:
1 = stopped normally, 0 = stopped manually
with [Quit] or [Stop].