Metrohm tiamo 2.2 Manual User Manual
Page 781
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.2
■■■■■■■■
767
Max. rate
Rate at which dosing is carried out outside of the Control range. The max-
imum dosing rate depends on the cylinder volume of the exchange unit or
dosing unit used.
Titrando
Range
0.01 ... 166.00 mL/min
Default value
1.00 (slow), maximum (optimal), maximum
(fast) mL/min
Titrino
Range
0.01 ... 150.00 mL/min
Default value
1.00 (slow), maximum (optimal), maximum
(fast) mL/min
Min. volume increment
This parameter determines the volume increment which is added at the start
and finish of the titration. The smaller the increment that is selected, the
slower the titration and the higher the accuracy.
Range
0.1 ... 9.9 µL
Default value
minimum (slow), minimum (optimal), 5.0
(fast) µL
Stop criterion
Stop criterion
The titration us stopped after the last dosing once the endpoint has been
reached and the stop criterion has been fulfilled. The titration can be stopped
when a particular drift has been achieved or after a preset time.
Titrino
Selection
drift | time
Default value
drift
Titrando
Selection
drift | rel. drift | off | time
Default value
drift
drift
The titration is stopped when the stop drift has been reached.
rel. drift
The titration is stopped when the relative stop drift (drift value at time of
start + relative stop drift) has been reached.
off
The titration will not be stopped.