5 kft ipol - titration parameters, Titration parameters – Metrohm tiamo 2.2 Manual User Manual
Page 762
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
748
■■■■■■■■
tiamo 2.2
Relative stop drift
The titration is stopped when the endpoint and the stop drift have been
reached. The stop drift is the sum of the drift at the titration start and the
relative stop drift defined here. This parameter is only shown for Stop cri-
terion = rel. drift.
Titrando
Range
1 ... 999 µL/min
Default value
10 µL/min
Delay time
If the endpoint has been reached, the delay time defined here is awaited
after the last dosing and then the titration is stopped. This parameter is only
shown for Stop criterion = time. The selection inf. stands for infinite.
Titrando, 855
Range
0 ... 999 s
Default value
10 s
Titrino
Range
0 ... 999 s
Default value
10 s
Selection
inf.
Stop time
The titration is stopped if the absolute stop time defined here has elapsed
since the start of the titration. This parameter is only shown for Stop cri-
terion = time and Delay time = inf.
Titrino
Range
1 ... 999999 s
5.6.3.5.2.5
KFT Ipol - Titration parameters
Tab: Method
▶ KFT Ipol ▶ Properties... ▶ Titration parameters
Command name
Name of the command.
Input
25 characters
The parameters for the running of the titration are defined on this tab.
Titration direction
Selection of the titration direction.
Selection
+ | - | auto
Default value
auto