5 Method
tiamo 2.2
■■■■■■■■
631
Initial measured value
Note
Is displayed only for Titrando and 855.
The initial measured value is determined before the other start conditions
are processed. The measured value acceptance depends on the three fol-
lowing parameters:
Messwertdrift
Der Messwert wird erst übernommen, wenn die Messwertdrift den hier ein-
gegebenen Wert unterschreitet.
Bereich
0.01 ... 99.90 µA/min
Auswahl
aus
Standardwert
aus
aus
Die Messwerte werden erst nach der maximalen Wartezeit übernommen.
Min. waiting time
The minimum waiting period is only important for drift-controlled measure-
ments. The measured value is not accepted until the minimum waiting time
has elapsed, even if the signal drift has already been reached. The drift con-
tinues to be checked while the waiting time is elapsing.
Range
0 ... 999999 s
Default value
0 s
Max. waiting time
If signal drift has been switched off or has not yet been reached, then the
measured value will be accepted when the maximum waiting time has
elapsed.
Range
0 ... 999999 s
Default value
1 s
Start volume
Start volume
Volume to be added before the start of the titration at the dosing rate indi-
cated.
Titrando, 855
Range
0.00000 ... 9999.99 mL
Default value
0.00000 mL