5 Method
tiamo 2.2
■■■■■■■■
1125
Target
Tower
Selection of the tower of the Sample Processor for moving to the required
lift position. Both Tower 1 and Tower 2 can always be selected even if the
Sample Processor has only one tower.
Selection
1 | 2
Default value
1
Lift position
Entering the desired lift position in mm or selecting a lift position predefined
for the rack Work position, Shift position, Rinse position or Special
position. A lift height of 0 mm corresponds to the Home position. Here
the lift is moved to the uppermost stop.
730, 778, 789, 814, 815, 855, 864
Range
0 ... 235 mm
Selection
Work position | Shift position | Rinse position |
Special position | Home position
Default value
Work position
774
Range
0 ... 100 mm
Selection
Work position | Shift position | Rinse position |
Special position | Home position
Default value
Work position
874
Range
0 ... 110 mm
Selection
Work position | Shift position | Rinse position |
Special position | Home position
Default value
Work position
Note
Please make sure that none of the lift positions exceed the maximum
stroke path given in the tower properties of the device.
Parameters
Lift rate
Selection of the speed for moving the lift.